Citizen CBM-910 Type II Instrukcja Obsługi Strona 99

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 114
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 98
Chapter 8 Periodic Inspection Mode
F-72 87
8.2 ION Periodic Inspection Mode Setting
In the ION periodical check mode, the instrument operations are checked by inputting the
output of the optional checker X-51 into the instrument.
Refer also to the instruction manual of the checker X-51 before the operation.
Set the measurement unit for ion to mol/L before starting the check operation.
After the setting change, the settings apply.
Follow the guidance to calibrate the instrument by simulated input of 0 mV and 120 mV in
the order below and check the following operation.
Simulated calibration check
0 mV 1.00 mol/L
120 mV 10.0 mmol/L
Simulated linearity check
240 mV 100 µmol/L
120 mV 10.0 mmol/L
0 mV 1.00 mol/L
120 mV 100 mol/L
Simulated temperature check
0.0°C
30.0°C
60.0°C
100.0°C
When all the check is completed, the result is outputted automatically.
Przeglądanie stron 98
1 2 ... 94 95 96 97 98 99 100 101 102 103 104 ... 113 114

Komentarze do niniejszej Instrukcji

Brak uwag